Performance Comparison of Point-of-Care Testing (Glucometer) and Laboratory Reference Glucose Oxidase Peroxidase (GOD-POD) Method for Glucose Measurement in Neonatal Jaundice
Navin Satyanarayan
Published : March 31, 2019
Abstract
Objective: This retrograde study was designed to compare between POCT glucometer and Laboratory reference method for glucose measurement in neonates especially in Neonatal Jaundice. The main objective of the study was to find the acceptability of POCT Glucometer against laboratory reference method. Material and Methods: 200 samples data were collected from the laboratory of neonatal jaundice. The samples neonatal jaundice with total Bilirubin >2mg/dL was considered for analysis., Glucose dehydrogenase method(Glucometer) and GOD-POD method (Reference Laboratory method) was used to measure glucose measurement in samples that were obtained. Results: Glucometer accuracy was evaluated using linear regression, Passing-Bablock regression, Bland-Altman analysis. There was no significant difference. Clarke Error Grid analysis, >98% results were in zone A. The Mean bias of Glucometer was 1.9%; with P <0.05. Conclusion: POCT Glucometer (Glucose dehydrogenase method) measurement performance was acceptable in hypoglycemic range especially in Neonatal Jaundice.